    Care After Use: Air Scaler

    1. Preparation

    • Remove the air scaler from the coupling joint and detach the tip.
      Bring air scalers to a well-prepared cleaning room without contamination.
      Remove organic contaminants with a paper wipe.

    2. Cleaning

    • Manual: Clean the external surface of air scaler with running water
      ( Automated: desinfection-ready-symbol NSK air scalers with this symbol are the thermo-disinfector-ready
      and may be washed and disinfected using a thermo-desinfector.
      When using a thermo-desinfector, be sure to follow its manufacturer's instructions.

    3. Desinfection

    • Manual: Wipe the external surface of air scalers with cleaning or alcohol for disinfection.

    4. Lubrication

      • Essential after every thermodisinfection process
      • Essential prior to each sterilisation process

      Manual: Supply 1-2 drops of NSK Turbine Oil to the drive air port of the air scalers. Connect to the coupling joint and operate for 5 seconds.
      Note: Excessive oil may cause instability in instrument rotation. Please consult the operation manual before use.

    5. Sterilisation

    • Place air scalers in a sterilisation pouch and seal them.
      EN13060 4.6.3 recommends autoclaving for 3 minutes (minimum holding time) at 134℃ or 15 minutes (minimum holding time) at 121℃ .
      NSK recommends Class B or S sterilisation.

    6. Storage

    • Immediately after a sterilisation cycle, remove air scalers from the autoclave.
      Keep them in a Dust-free sterilised storage or bring them to a treatment room for use.